Appendix C
Alarm Testing Procedure
REF 1057983 A Respironics V200 Ventilator Operator’s Manual Appendix C-3
4. Alarm should reset. (If the backup battery is attached, the ventilator
should return to AC power. If no battery is attached, the ventilator will
resume operating.)
Apnea Alarm Test
1. Set the rate to 1 breath per minute.
2. Wait for 20 seconds.
3. The Respironics V200 should begin ventilating at a rate of 20 BPM
while activating the Apnea alarm.
4. Reset the rate to 10 BPM.
5. RESET the alarm.
6. The Respironics V200 should begin ventilating normally.