13-104 Respironics V200 Ventilator Operator’s Manual REF 1057983 A
Options and Accessories
Do not use with neonatal tracheostomy tubes.
Do not use on neonates.
Do not inflate the tracheostomy tube cuff until Speaking Mode has
been turned off.
Do not use on patients who are dependent on PEEP therapy.
Do not use with HME filters.
Discontinue using Speaking Mode promptly and institute appropriate
ventilation if the patient experiences difficulty or their status
deteriorates from baseline parameters.
Perform an airway patency test immediately after activating Speaking
Mode.
NOTE: Speaking Mode is available ONLY in invasive ventilation mode.
Patient Preparation Using Speaking Mode on a patient requires preparation before activating the
mode. Refer to the Sequence for Use diagram in Figure 13-58 and follow the
necessary steps before starting Speaking Mode and discontinuing Speaking
Mode. The Speaking Mode Clinical Guide describes patient care steps in
detail.
WARNING: Read and understand these Speaking Mode Option instructions before using
Speaking Mode on a patient.
WARNING: Wait at least 72 hours after a tracheostomy before initiating Speaking
Mode.