Chapter 10
10-8 Respironics V200 Ventilator Operator’s Manual REF 1057983 A
Care and Maintenance
Table 10-1: Schedule for Periodic Maintenance
Removing Cooling Filter
Figure 10-4: Removing Cooling Filter
Refer to Figure 10-4.
1. Remove black filter retaining bracket (1). (Pry off if necessary.)
2. Wash filter (2) in soapy water.
3. Dry thoroughly before reinstalling.
12,500 hours • 12,500 hour preventive
maintenance kit (P/N
1001733). Kit contents are
subject to change.
• Install 12,500 hour preventive
maintenance kit
• Clean ventilator interior and exterior
• Complete performance verification
procedure
CAUTION: The 12,500 hour preventive maintenance procedure is to be
performed only by a qualified service technician.
As required • External oxygen sensor
• Backup Battery
• Replace and recalibrate new sensor by
running Extended Self-Test
• For charging and maintenance
instructions, see “Backup Battery” on
page 4-4.
CAUTION: Do not remove any screws from the cooling filter area. Removing screws
from this area will result in damage to internal components.
1
2