Philips V200 Ventilation Hood User Manual


 
Chapter 12
Technical Specifications
REF 1057983 A Respironics V200 Ventilator Operator’s Manual 12-11
Leakage Current
Compliance and
Approvals
Power Requirements
Leakage Current
Ventilator Earth Leakage Current 100 to 240VAC; 300µA maximum
Enclosure/Patient Leakage Current 100 to 240VAC; 100µA maximum
Table 12-22: Leakage Current
WARNING: When connecting a humidifier to the humidifier outlet (available only on
100-120 VAC ventilators) allowable leakage current values may be
exceeded.
Compliance and Approvals
The Respironics V200 ventilator system complies with the
requirements of the European Directive 93/42/EEC
concerning medical devices and the requirements of Directive
89/336/EEC relating to electromagnetic compatibility.
IEC 601-1 Classification Protection class 1, type B applied part, drip proof, continuous
operation.
The ventilator system complies
with these international and
European Standards.
IEC 601-1/EN60601-1
IEC 601-1-2/EN60601-1-2
EN 794-1
The ventilator system has been
certified by the following.
CSA/NRTL: CSA C22.2 No. 601-1, CSA C22.2 No. 601-2-
12, UL2601-1
Table 12-23: Compliance and Approvals
Power Requirements
Configuration Voltages & Frequencies
Ventilator Only 100 to 240VAC, 50/60Hz, 6 amp max. or
100 to 120VAC, 50/60Hz, 6 amp max. depending on
configuration
Ventilator Only 24VDC, 8AH, Backup Battery pack (operating time
approximately 30 minutes)
Ventilator with Humidifier 100 to 120VAC, 50/60Hz, 9 amp max.
Table 12-24: Power Requirements