Emerson Process Management 3-9000-745 Wok User Manual


 
User Manual MON20/20 Software for Gas Chromatographs
3-9000-745 SEPTEMBER 2010
3-8
Figure 3-6. The Heaters window
2. For each heater that you want to set, select Fixed On from the
appropriate row under the Switch column.
3. For each heater that you want to set, double-click on the appropriate
row under the Fixed PWM Output column, and enter the desired
percentage of output. You can enter a decimal value between 0 and
100.
4. To save the changes and leave the window open so that you can
monitor the heaters’ status, click Save. The current temperature of
each heater displays in the Temperature column, and is updated in
real time.
5. To save the changes and close the window, click OK.
It is not recommended that a value of 95 or higher be used for a prolonged time, as this
may damage the equipment.
CAUTION