MON20/20 Software for Gas Chromatographs User Manual
SEPTEMBER 2010 3-9000-745
4-17
The following data displays for each peak:
3. Click Close to return to the component data table.
4.3 Managing timed events
Use this function to view and/or edit the timed events tables assigned to
and used by particular gas streams. The number of available timed
events depends on the GC unit configuration. The standard GC
application contains four timed events tables.
Note
See “Editing Timed Events from the Time Events window” on page 2-33 for more
information.
Name Description
Peak No Numerical identifier for the peak, listed by the order of discovery.
Ret Time Time, in seconds, that the component eluted.
Peak Area The area under the peak.
Peak Height The maximum height of the peak.
Det # The detector associated with the peak.
Method Method of peak end detection. Options are:
1 (Baseline)
2 (Fused Peak)
3 (Last Fused Peak)
•4 (Tangent Skim)
100 (Inhibit)
300 (Forced Integration)
500 (Summation)
Integration Start Time, in seconds, when integration started.
Integration Stop Time, in seconds, when integration stopped.
Peak Width Half Height The width of the peak taken at half of the peak’s height.
Is Partial Peak If Y, then the Partial Peak value is used in the summation calculation; if N,
then the Partial Peak value is not used in the summation calculation.