SEPTEMBER 2010 3-9000-745
4-31
3. Click on the Analysis Time cell and enter a value, in seconds, between
0 and 3600.
4. Click on the Cycle Time cell and enter a value, in seconds, between 0
and 3620.
Note
The Cycle Time must be atleast 10 seconds greater than the Analysis Time.
5. To save the changes without closing the window, click Save. To save
the changes and close the window, click OK.
4.3.5 Removing an event from the Timed Event Table
To remove an event from one of the Valve Events, Integrate Events, or
Spectrum Gain Events tables on the Timed Events window, do the
following:
1. Select Timed Events... from the Application menu. The Timed
Events Tables selector window appears, displaying a list of available
timed events tables.
Figure 4-23. The Timed Events Tables selector window
Note
Other ways of accessing the timed event tables are by pressing F5 or by clicking
from the Toolbar.