Emerson Process Management 3-9000-745 Wok User Manual


 
User Manual MON20/20 Software for Gas Chromatographs
3-9000-745 SEPTEMBER 2010
3-50
5. Enter the Zero Scale Adjustment value and then click Next. If the
value entered is within tolerance, it is accepted and Step 3 of the
Analog Output Calibration Wizard displays. If the value is not within
tolerance, an error icon ( ) appears beside the field. Tolerance is set
to ±1mA of the analog output’s default zero adjustment setting, which
is 4mA. Enter a different value and try again.
Figure 3-43. Step 3 of the Analog Output Calibration Wizard
6. Enter the Full Scale Adjustment value and then click Next. If the
value entered is within tolerance, it is accepted and Step 4 of the
Analog Output Calibration Wizard displays. If the value is not within
tolerance, an error icon ( ) appears beside the field. Tolerance is set
to ±1mA of the analog output’s default full adjustment setting, which
is 20mA. Enter a different value and try again.