Emerson Process Management 3-9000-745 Wok User Manual


 
User Manual MON20/20 Software for Gas Chromatographs
3-9000-745 SEPTEMBER 2010
3-24
4. To raise the baseline, click Right(N). Each time this button is
clicked, N is incremented by 1. For example, is this is the first time
the button has been clicked, Right(0) will be increment to Right(1) and
the baseline will be raised one step. If Left(N) was clicked previously,
then that button will be incremented by 1 first, until it reaches Left(0);
at the point, Right(N) will be incremented by 1.
Note
To reset the baseline to its original setting, click Right(N) and Left(N) until they read
Right(0) and Left(0).
3.4 Managing your gas chromatograph’s discrete inputs
You can use MON 20/20 to assign labels to the GC’s discrete inputs and to
control the discrete inputs’ operational modes. The number of discrete
inputs available depends on the GC.
Note
This window contains a hidden column labelled Physical Name. For more information
about this column and how to display it, see “Viewing the Physical Name column” on
page 1-38.
3.4.1 Renaming a discrete input
Give each discrete input a descriptive label to avoid confusing one unit for
another. To assign an identifying label, do the following:
1. Select Discrete Inputs... from the Hardware menu. The Discrete
Inputs window displays.