Emerson Process Management 3-9000-745 Wok User Manual


 
MON20/20 Software for Gas Chromatographs User Manual
OCTOBER 2010 3-9000-745
C-15
9009 1 integer 2 1 system time hour (0-23)
9010 1 integer 2 1 system time minutes (0-59)
9011 1 integer 2 1 system time seconds (0-59)
9012 1 integer 2 1 system time day (0-6)
9013 0 integer 2 1 plug ID (Modbus or Device ID, per DIP switch settings)
9014 1 long 4 2 site ID
9016 0 string 12 5 device serial number
9022 0 integer 2 1 analysis time 1
9023 0 integer 2 1 analysis time 2 (for dual detector system)
9024 0 integer 2 1 cycle time 1
9025 0 integer 2 1 cycle time 2 (for dual detector system)
9026 0 integer 2 1 run time 1
9027 0 integer 2 1 run time 2 (for dual detector system)
9028 0 integer 2 1 current stream 1
9029 0 integer 2 1 current stream 2 (for dual detector system)
9030 0 integer 2 1 system mode 1
9031 0 integer 2 1 system mode 2 (for dual detector system)
9032 0 integer 2 1 calibrating 1
9033 0 integer 2 1 calibrating 2 (for dual detector system)
9034 0 integer 2 1 active alarm (red light at GC controller)
9035 0 integer 2 1 unack’d alarm (yellow light at GC controller)
9036 0 integer 2 1 hourly average reset - year
9037 0 integer 2 1 hourly average reset - month
9038 0 integer 2 1 hourly average reset - day
9039 0 integer 2 1 hourly average reset - hour
9040 0 integer 2 1 hourly average reset - minutes
9041 0 integer 2 1 24-hour average reset - year
9042 0 integer 2 1 24-hour average reset - month
9043 0 integer 2 1 24-hour average reset - day
Table C-7 SIM_2251 Modbus Communication Indices
Reg.
No.
Description: RW = read/write (1) or read-only (0)
LEN = length
REGS = number of Modbus registers required
RW TYPE LEN REGS VARIABLE NAME, POINTER, OR DESCRIPTION