Emerson Process Management 3-9000-745 Wok User Manual


 
MON20/20 Software for Gas Chromatographs User Manual
SEPTEMBER 2010 3-9000-745
3-29
3.5 Managing your gas chromatograph’s discrete outputs
You can use MON 20/20 to assign labels to the GC’s discrete outputs and
to control the discrete outputs’ operational modes. The number of discrete
outputs available depends on the GC.
Note
This window contains a hidden column labelled Physical Name. For more information
about this column and how to display it, see “Viewing the Physical Name column” on
page 1-38.
3.5.1 Renaming a discrete output
Give each discrete output a descriptive label to avoid confusing one unit
for another. To assign an identifying label, do the following:
1. Select Discrete Outputs... from the Hardware menu. The Discrete
Outputs window displays.
Figure 3-20. The Discrete Outputs window
2. Double-click on the appropriate row under the Label column for the
discrete output that you want to rename.