Emerson Process Management 3-9000-745 Wok User Manual


 
MON20/20 Software for Gas Chromatographs User Manual
SEPTEMBER 2010 3-9000-745
4-79
Figure 4-69. The Stream Sequence window
2. Each stream sequence table can contain up to three sequences--a
primary, or default, sequence, and two auxiliary sequences. The table
for detector that is designated as “1” displays by default. To display a
different table, select it from the Choose table: drop-down list.
3. To create a new stream sequence, click Insert before.
Note
There can only be three sequences per detector. If a detector already has three
sequences and you want to create a new one, you must edit or delete one of the existing
sequences. Click Delete to delete a sequence.
4. Double-click the appropriate cell under the Strm Seq Name column to
give your new sequence a name, or to edit the name of an existing
sequence. Type in the new name.
5. To define the order of analysis, double-click the appropriate cell under
the Seq of Strms column and the numbers for the streams, seperated
by commas, that should be analyzed.
6. To define which discrete input should activate the sequence, select it
from the drop-down list of the appropriate cell under the Seq Activate
DI column.