Emerson Process Management 3-9000-745 Wok User Manual


 
MON20/20 Software for Gas Chromatographs User Manual
SEPTEMBER 2010 3-9000-745
4-77
Details about the valves in the drop-down list can be viewed from the
Valves window.
3. To save the changes without closing the window, click Save. To save
the changes and close the window, click OK.
4. To have the stream open when the valve is ON, select the
corresponding Stream Valve On to Select checkbox; to have the stream
open when the valve is OFF, clear the corresponding Stream Valve On
to Select checkbox.
4.10.3 Assigning a data table to a particular stream
To assign a component data table, a validation data table, or a timed
events table to a stream, do the following:
1. Select Streams... from the Application menu. The Streams window
opens.
Figure 4-67. The Streams window
2. For the appropriate stream, if Usage is set to Cal or Analy, select a
component data table from the CDT column and a timed events table
from the TEV column.
3. For the appropriate stream, if Usage is set to Validate, select a
component data table from the CDT column, a timed events table from
the TEV column, and a validation data table from the VDT column.