SEPTEMBER 2010 3-9000-745
4-29
The following table describes the spectrum gain-related parameters that
are available on the timed events window.
4.3.4 Setting the cycle and analysis time
To set the cycle and analysis time, do the following:
1. Select Timed Events... from the Application menu. The Timed
Events Tables selector window appears, displaying a list of available
timed events tables.
Figure 4-21. The Timed Events Tables selector window
Note
Other ways of accessing the timed event tables are by pressing F5 or by clicking
from the Toolbar.
Parameter Description
Det # The ID number of the detector that will be affected by the event. Select 1 or 2.
Gain Enter a value between 0 and 64. This is the exponent value in the following
expression: 2
gain value
. For example, a value of 0 means no gain is applied; a
value of 5 means the gain is increased to 32 times it’s original value.
Time Indicates the time, in seconds, that the event should occur during the
analysis. Enter a value between 0.0 and 3600.0.
NOTE: Event times must be less than the analysis time.