MON20/20 Software for Gas Chromatographs User Manual
SEPTEMBER 2010 3-9000-745
3-25
Figure 3-16. The Discrete Inputs window
2. Double-click on the appropriate row under the Label column for the
discrete input that you want to rename.
Note
The discrete inputs are labelled Discrete Input 1 - Discrete Input N by default,
where N equals the total number of discrete inputs available to the GC.
3. Type in a new descriptive name for the discrete input.
4. Click OK.
3.4.2 Setting a discrete input’s operational mode
A discrete input has three operational modes: Auto, On, and Off.
• Setting the discrete input to Off means that it will interpret all
incoming signals as OFF, despite the true nature of the signal.
• Setting the discrete input to Auto means that it will analyze the
incoming signal to determine whether it is ON or OFF.
• Setting the discrete input to On means that it will interpret all
incoming signals as ON, despite the true nature of the signal.