Emerson Process Management 3-9000-745 Wok User Manual


 
User Manual MON20/20 Software for Gas Chromatographs
3-9000-745 OCTOBER 2010
C-14
7092 Compressibilty (Last Calibration)
7093 Total GPM (Last Calibration)
7094 Total Unnormalized (Last Calibration)
7095-7110 Response Factors (#1-16), Component Table n (where n equals the CDT # used during
the last run)
7111-7126 Response Factors (#1-16), Component Table n (where n equals the CDT # used during
the last run)
7127-7162 Averages, User-define 1-36
Note: Registers 7127-7141 are copies of registers 7055-7069.
7163-7198 Maximum values from averages, User-define 1-36
7199-7234 Minimum values from averages, User-define 1-36
7235-7270 First (most recent) archive of averages, User-define 1-36
Note: Registers 7235-7249 are copies of registers 7070-7084.
7271-7306 First (most recent) archive of maximum values from averages, User-define 1-36
7307-7342 First (most recent) archive of minimum values from averages, User-define 1-36
7343-7378 Second archive of averages, User-define 1-36
7379-7414 Second archive of maximum values from averages, User-define 1-36
7415-7450 Second archive of minimum values from averages, User-define 1-36
7451-7486 Third (oldest) archive of averages, User-define 1-36
7487-7522 Third (oldest) archive of maximum values from averages, User-define 1-36
7523-7558 Third (oldest) archive of minimum values from averages, User-define 1-36
Table C-7 SIM_2251 Modbus Communication Indices
Reg.
No.
Description: RW = read/write (1) or read-only (0)
LEN = length
REGS = number of Modbus registers required
RW TYPE LEN REGS VARIABLE NAME, POINTER, OR DESCRIPTION
9001 0 string 6 3 device model number
9004 0 string 4 2 software revision
9006 1 integer 2 1 system time month (1-12)
9007 1 integer 2 1 system time day (1-31)
9008 1 integer 2 1 system time year (0-99)
Table C-6 SIM_2251 Modbus register list (floating point)
Reg. No. Description