Emerson Process Management 3-9000-745 Wok User Manual


 
User Manual MON20/20 Software for Gas Chromatographs
3-9000-745 SEPTEMBER 2010
5-22
5.6 Displaying reports
This function allows you to immediately display, print, or store
preconfigured reports of GC analysis data. Data is reported in real-time
from the GC or from saved files.
5.6.1 Understanding report types
MON 20/20 can generate the following types of reports:
Analysis: Displays a list of the components that were detected, based
on raw data. Displays a list of calculations for each component, based
on the table located at Application Calculations Control....
See “Setting standard calculations by stream” on page 4-37 for more
information.
There are two types of analysis reports: Analysis (GPA) and Analysis
(ISO). See page 5-25 for an example Analysis (GPA) report. See
page 5-26 for an example Analysis (ISO) report.
Calibration: Displays a list of the components that were detected,
along with each component’s calibration concentration, raw data
value, new response factor, and new retention time. See page 5-27 for
an example report.
Final Calibration: The Final Calibration report displays the list of
components along with each component’s old and new response
factors, and each component’s old and new retention times, based on
the averaged data. See page 5-28 for an example report.
Validation: For the most recent validation cycle, displays the
Nominal Value, Allowed Percent Deviation, and the Measured Value
of each variable in the Validation Data table. See page 5-29 for an
example report.
Note
If the actual deviation is beyond the allowed amount, then the row will be flagged with
an *.