SEPTEMBER 2010 3-9000-745
4-99
Note
The PV Status column indicates the state of the data displayed in the PV Value column.
If the data was generated under predictable conditions, then the status for all mapped
process variables will be Good; if the data was generated under unpredictable
conditions—that is, if any alerts were triggered during the analysis cycle—then the
status for all mapped process variables will be Bad, because the GC cannot guarantee
the results of the analysis.
3. To copy the stream settings from a highlighted row and apply them to
the next row, click S + Copy. This feature also increments the Stream
value to the next available stream—for instance, incrementing from
Stream 2 to Stream 8, per the GC application.
4. To copy the component settings from a highlighted row and apply
them to the next row, click C + Copy. This feature also increments the
Component value to the next available component—incrementing
from Ammonia to Benzene, per the GC application.
5. If necessary, enter a date or time format into the Date/Time Format
column.
6. To save the changes without closing the window, click Save. To save
the changes and close the window, click OK.