MON20/20 Software for Gas Chromatographs User Manual
SEPTEMBER 2010 3-9000-745
4-71
4. If you want a discrete output to activate when the alarm triggers, click
on the appropriate cell under the DO # to Set column and select it
from the drop-down list.
5. To prevent or allow averaging when the alarm triggers, double-click
on the appropriate cell under the Inhibit Avg column, and select one of
the following options:
True - Inhibits averaging when the alarm is active.
False - Allows averaging when the alarm is active.
6. To customize the text of the alarm message, enter the new text in the
appropriate cell under the User Alarm Text column. When the alarm
triggers, this text will display under the Alarm Message column on the
Unack/Active Alarms window.
Note
If an alarm message is changed, all affected alarm entries, including those previously
recorded, will include that change.
7. To enable or disable the use of the customized alarm text, select True
or False from the appropriate cell under the Inhibit Alarm Text
column.
8. To copy the stream settings from a highlighted row and apply them to
the next row, click S + Copy. The stream will be copied and
incremented to the next available stream--for example, from Stream 2
to Stream 3.
If there are no streams available, instead of copying the stream, MON
20/20 will display the following message:
Figure 4-62. No streams available warning