MON20/20 Software for Gas Chromatographs User Manual
SEPTEMBER 2010 3-9000-745
4-75
Figure 4-65. The Streams window
2. For the appropriate stream, select one of the following options from
the Usage column:
Unused - Not used
Cal - Calibration
Analy - Analysis
Validate - Validation
3. If you select Cal or Validation, you can also edit the following
parameters:
Auto - If checked, the calibration or validation will be automatic.
Tot Runs - The number of runs, from 1 to 10, to make for each
calibration.
Avg Runs - The number of most-recent calibration runs to
average; for instance, if five calibration runs are performed and
Avg Run is set to 3, then the last three runs of the five will be used
to average the calibration results.
Start Time - The time the first automatic calibration should be
performed.
Interval - The number of hours between automatic calibrations.
Auto Calib - Enable or disable the automatic calibration run.