Emerson Process Management 3-9000-745 Wok User Manual


 
MON20/20 Software for Gas Chromatographs User Manual
SEPTEMBER 2010 3-9000-745
3-21
4. To raise the baseline, click Right(N). Each time this button is
clicked, N is incremented by 1. For example, is this is the first time
the button has been clicked, Right(0) will be increment to Right(1) and
the baseline will be raised one step. If Left(N) was clicked previously,
then that button will be incremented by 1 first, until it reaches Left(0);
at the point, Right(N) will be incremented by 1.
Note
To reset the baseline to its default setting, click Right(N) and Left(N) until they read
Right(0) and Left(0).
5. After the baseline has been raised or lowered to your satisfaction, click
OK.
3.3.2 Igniting the FID flame
If the FID Ignition field on the Detectors window is set to “Manual” and if
the Flame status field is set to “Off”, do the following to restart the flame:
1. Click Open H2 Valve. The H2 Valve Cur State field changes to
“Open”.
2. Click Ignite. The Flame Status field changes to “On” when the FID
internal temperature exceeds the value set in the Flame On Sense
Temp field.
Note
If the FID Ignition field is set to “Auto”, the GC will automatically restart the flame if it
goes out.