Emerson Process Management 3-9000-745 Wok User Manual


 
MON20/20 Software for Gas Chromatographs User Manual
SEPTEMBER 2010 3-9000-745
3-43
3.7 Managing your gas chromatograph’s analog outputs
With MON 20/20 you can control them in the following ways:
Assign identifying labels.
Assign scale ranges.
Calibrate analog outputs for zero and full scale values.
Note
This window contains a hidden column labelled Physical Name. For more information
about this column and how to display it, see “Viewing the Physical Name column” on
page 1-38.
3.7.1 Renaming an analog output
Give each analog output a descriptive label to avoid confusing one unit for
another. To assign an identifying label, do the following:
10.Select Analog Outputs... from the Hardware menu. The Analog
Outputs window displays.
Figure 3-34. The Analog Outputs window
11.Double-click on the appropriate row under the Label column for the
analog output that you want to rename.