MON20/20 Software for Gas Chromatographs User Manual
OCTOBER 2010 3-9000-745
2-17
Figure 2-13. The Edit Scales window
The following table lists the parameters that can be edited:
Command Description
Default
Value
X Min Sets the minimum value, in seconds, for the X axis. 0
X Max Sets the maximum value, in seconds, for the X axis. The is
value is determined by the Timed Events table.
100
Y Min Sets the minimum value for the Y axis. -10
Y Max Sets the maximum value for the Y axis. 100
Print Speed Sets the number of inches per second for the x-axis while
printing a chromatogram, similar to an XY plotter.
0
X Intervals Sets the number of intervals to be displayed on the graph for
the X axis.
10
Y Intervals Sets the number of intervals to be displayed on the graph for
the Y axis.
11
Display Option Determines whether the chromatograph is displayed as a
solid line or as a dotted line. Lines is checked by default.
Lines
Show labels Toggles the display of the graph labels. Checked
Scroll newest X Determines whether the graph’s window moves to focus on
the most recent data point along the x axis. This feature
only applies to live chromatograms.
Unchecked