Auto Baseline - Enable or disable the automatic baseline run.
The GC performs an additional calibration run (before the
calibration runs to be averaged) without the calibration gas. This
run evaluates the peaks caused by the GC valve action alone; any
peak areas found are subtracted from the subsequent analyses.
Note
Disabling the Auto Baseline setting will delete existing CDT baseline data for the
associated stream.
4. To save the changes without closing the window, click Save. To save
the changes and close the window, click OK.
4.10.2 Assigning a valve to a stream and setting the relationship
between the stream’s open state to the valve’s On/Off state
To assign a valve to a stream, do the following:
1. Select Streams... from the Application menu. The Streams window
opens.
Figure 4-66. The Streams window
2. Go to the Stream Valve column for the corresponding stream and
select the appropriate valve from the drop-down list.