Emerson Process Management 3-9000-745 Wok User Manual


 
User Manual MON20/20 Software for Gas Chromatographs
3-9000-745 OCTOBER 2010
B-4
B.3.1 Concentration analysis with response factor
Calibration
The concentration calculations discussed as follows require a unique
response factor for each component in an analysis. These factors may be
manually entered by an operator or automatically calculated by
calibrating the system.
where
Calculated response factors are stored by the GC for use in the
concentration calculations, and are printed out in the configuration and
calibration reports.
or
ARF
n
Area response factor for component n in area per mole percent (%).
HRF
n
Height response factor for component n.
Area
n
Area associated with component n in calibration gas.
Ht
n
Height associated with component n in mole percent in calibration gas.
Cal
n
Amount of component n in mole percent of calibration gas.
ARF
n
Area
n
Cal
n
---------------
= HRF
n
Ht
n
Cal
n
-----------
=
RFAVG
n
RF
i
i 1=
k
k
------------------
=