Emerson Process Management 3-9000-745 Wok User Manual


 
User Manual MON20/20 Software for Gas Chromatographs
3-9000-745 SEPTEMBER 2010
3-48
The operational status of each analog output displays under the Status
column. There are three possible status readings, and their meanings are
as follows:
This window also displays other types of data, such as the following:
mA - The amount of current being generated in milliamperes.
Cur Val - The current scaled value of the analog output signal.
3.7.6 Calibrating an analog output
To automatically calibrate an analog output, do the following:
1. Select Analog Output... from the Hardware menu. The Analog
Outputs window displays.
Figure 3-40. The Analog Outputs window
2. Click on the analog output that you want to calibrate.
OK The analog output device is installed and is working
correctly.
Not Installed The analog output device is not installed.
Error The Heater/Solenoid board is installed but the GC cannot
communicate with it.