Emerson 300510EN Yogurt Maker User Manual


 
System Configuration Manual
300510EN, Rev AA
December 2010
5-69
Rosemount Raptor
Section 5. Installing a Raptor Level Gauging System
Manual Configuration
1. Ensure that the Use Auto Sensor Configuration check box is
unchecked.
2. Choose conversion method in the Method input field.
For more information on conversion methods such as User defined table,
User defined formula and User defined individual formula, see the
appropriate sections in “User Defined Temperature Conversion” on
page 7-4.
NOTE
For Spot and Multiple Spot Temperature elements, use the Spot Pt100 sensor
option in order to obtain a correct average temperature calculation.
3. Specify the measurement range of the current temperature element by
specifying the minimum and maximum temperatures in the Min Value
and Max Value input fields.
4. Choose the type of sensor connection that is used for the spot sensors in
the Connection drop-down list.
5. Click the Apply button to store the configuration, or click the OK button
to store the configuration and close the window.
See the Rosemount 2240S Multi-Input Temperature Transmitter Reference
Manual (Document No. 300550EN) for more information on how to configure
temperature sensors for the Rosemount 2240S.