7.10.4 Action Buttons The following buttons can be found in the Tank Scan window:
Save See “File Storage” on page 7-26.
Stored
Echoes
See “File Storage” on page 7-26.
Reread From
Gauge
At any time, you can refresh tank echo and echo
peaks with the Reread From Gauge button. The
Tank Scan will display the new echo curve as a black
line, and up to two previous echo curves in grey
color. The previous echo peaks will appear with
small crosses.
Print Opens the print dialogue and prints the Tank Scan
window.
Apply When changing a parameter that affects echo peak
detection (e.g. General Amplitude Threshold), you
will have to press the Apply button to write these
settings to the internal memory of gauge. It takes a
few seconds for the gauge to update the echo peak
data (up to 30 seconds due to the echo peak filtering
function in the gauge).
Finally, press the Reread from Gauge button to
update the echo peak information in the graph area.
OK Applies changes and closes the window.
Cancel Cancels all changes.
Help Opens the online help for the window.