Tektronix AWG710B Ventilation Hood User Manual


 
The Setup Menu Screen
AWG710&AWG710B Arbitrary Waveform Generator User Manual 3-35
Extended Operation Displays the Extended Operation side menu to select four operation
modes AWG mode, FG mode, Waveform Mixing mode, and Synchronous
Operation mode.
Table 3-4: Status area
Item Description
Clock Displays the clock rate of an internal clock or an external clock. The clock
rate of an external clock is measured and displayed when external clock
use. The following message is displayed in addition to a numerical value.
PLL UNLOCKED: There is no Ref In signal when Clock Src is
Internal and Clock Ref is External.
NO SIGNAL: There is no Clock signal when Clock Src is External.
OUT OF RANGE: The clock rate of a external clock signal is out of
range when Clock Src is External.
NOT STABLE: The clock rate of a external clock signal is within
range but not stable when Clock Src is External.
Run Mode Displays the Run Mode selected on the Run Mode menu.
Instrument statusl The following state of a sequencer of operation is displayed.
Stopped: Output operation has stopped.
Running: Output operation is executing.
Waiting: Waiting for a trigger signal or an event signal where the
run mode is except Continuous.
Table 3-3: Setup bottom menu buttons
Bottom menu
button Description