Appendix B: Performance Verification (AWG710B)
AWG710&AWG710B Arbitrary Waveform Generator User Manual B-53
b. Push SETUP (front–panel)!Extended Operation (bottom)!Sync -
Master (side).
c. Push Connect (bottom)!Execute Trig Timing Calibration (side).
3. Check that Pass is displayed being displayed on the Trigger Timing Cal. field
at the lower left of the AWG710B screen.
4. Change the cable connection.
a. Disconnect the cable from the T OUT1 connector and connect it to the T
OUT2 connector.
b. Disconnect the cable from the T OUT1
connector and connect it to the T
OUT2 connector.
c. Disconnect the cable from the C OUT1 connector and connect it to the T
OUT2 connector.
d. Disconnect the cable from the C OUT1
connector and connect it to the T
OUT2 connector.
5. Push Execute Trig Timing Calibration (side).
6. Check that Pass is displayed being displayed on the Trigger Timing Cal. field
at the lower left of the AWG710B screen.
7. Disconnect the four SMA cables.