Appendix B: Performance Verification (AWG710B)
B-6 AWG710&AWG710B Arbitrary Waveform Generator User Manual
Performance Tests
This section contains a collection of procedures for checking that the AWG710B
Arbitrary Waveform Generator performs as warranted.
The procedures are arranged in eight logical groupings, presented in the following
order:
The performance tests check all of the characteristics that are designated as
checked in Appendix A:Specifications. (The characteristic items that must be
checked are listed with the check mark (n) in Appendix A: Specifications).
Table B-1: Performance test items
Titles
See
(performance
verification) Test items See (specifications)
Operating mode tests Page B-13 Continuous, Triggered, and Gated
mode normality
Page A-2
Amplitude and offset accuracy tests (normal out) Page B-19 Amplitude accuracy, and DC offset
accuracy
Page A-4
Amplitude, DC offset and rise time accuracy tests
(direct DA out)
Page B-23 Amplitude accuracy, DC offset
accuracy and Rise time accuracy
Page A-4
Amplitude, DC offset and
Rise time accuracy test (Option2)
Page B-27 Amplitude accuracy, DC offset
accuracy and Rise time accuracy
Page A-4
Pulse response tests (normal out) Page B-31 Rise time accuracy Page A-4
Trigger input tests Page B-33 Trigger level accuracy, and Trigger
function normality
Event input and enhanced mode tests Page B-37 External event input function normality,
and
Event mode normality
External Clock input and VCO Out tests
(AWG710B only)
Page B-37 External Clock input function normality,
and VCO Out normality
Page A-6, A-18
VCO Out frequency and 10 MHz reference input
tests (AWG710B only)
Page B-48 VCO Out frequency and
Reference input normality accuracy
Page A-3
Marker output tests Page B-50 Marker output level accuracy Page A-5
Synchronous operation tests (AWG710B only) Page B-50 Synchronous operation normality Page A-5
NOTE. These procedures extend the confidence level provided by the basic
procedures described on page B-2. The basic procedures should be done first, and
then these procedures performed if desired.