Tektronix AWG710B Ventilation Hood User Manual


 
Appendix B: Performance Verification (AWG710)
AWG710&AWG710B Arbitrary Waveform Generator User Manual B-81
Amplitude, Offset Accuracy and Rise Time Tests (for option 02)
These procedures check the accuracy of the AWG710 Arbitrary Waveform
Generator option02 outputs; amplitude and offset.
Check Amplitude
and
DC Offset
Do the following steps to install the test hookup and set the test equipment controls:
1. Use a 50 BNC coaxial cable, an SMA(Ma)–BNC(Fe) adapter, a 50
precision terminator, and a BNC–to–dual banana adapter to connect the
AWG710 Arbitrary Waveform Generator CH1 output to the DMM input
connector (see Figure B-46).
Figure B-46: Option02 output amplitude accuracy initial test hookup
2. Set the DMM controls as follows:
3. Push UTILITY (front–panel)!System (bottom)!Factory Reset
(side)!OK (side).
Equipment
required
A 50 BNC coaxial cable, a 50 precision terminator, an
SMA(Ma)–BNC(Fe) adapter, BNC (female)–to–dual banana adapter, and
a digital multimeter (DMM).
Prerequisites
The AWG710 option 02 Arbitrary Waveform Generator must meet the
prerequisites listed on page B-61.
Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VDC
Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 V
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Front
AWG710 option 02 Arbitrary Waveform Generator
DMM
50 BNC coaxial cable
CH1 output connector
+ SMA(Male)-BNC(Female) adapter
+ BNC 50Ω precision terminator
+ 50 Ω BNC coaxial cable
Input connector
+ BNC-to-dual banana adapter
+ 50 BNC coaxial cable