Appendix B: Performance Verification (AWG710)
B-102 AWG710&AWG710B Arbitrary Waveform Generator User Manual
a. Disconnect the 50 Ω precision terminator from the MARKER1
OUT
connector.
b. Disconnect the cable (with SMA(Ma)-BNC(Fe) adapter and BNC 50 Ω
terminator) from MARKER 1 OUT connector.
c. Connect the 50 Ω precision terminator to the MARKER 1 OUT (removed
in step a).
d. Connect the adapter, terminator, and the cable (removed in step b) to the
MARKER1
OUT connector.
8. Press Force Event where needed.
9. Repeat steps 5 and 6 for the AWG710 Arbitrary Waveform Generator
MARKER1
OUT.
10. Repeat steps 5 through 8 for MARKER 2 OUT and MARKER2
OUT.
11. Disconnect the DMM.