Tektronix AWG710B Ventilation Hood User Manual


 
Appendix B: Performance Verification (AWG710)
B-92 AWG710&AWG710B Arbitrary Waveform Generator User Manual
3. Set the oscilloscope controls as follows:
4. Set all the switches of the ground closure to open.
5. Follow the substeps below to set the AWG710 Arbitrary Waveform Generator
controls and select the sequence file:
a. Push UTILITY (front–panel)!System (bottom)!Factory Reset
(side)!OK (side).
b. Load the PT_EVENT.SEQ file.
Refer to Loading Files on page B-62 for file loading procedures.
c. Push SETUP (front–panel)!Run Mode (bottom)!Enhanced (side) to
set the enhanced mode.
6. Push the RUN and CH1 OUT buttons.
The LEDs above the RUN button and CH1 output connector are on.
7. Check the EVENT IN connector pin 0 input:
a. Verify that a waveform displayed on the oscilloscope has the same
amplitude as shown in Figure B-55.
Vertical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CH1
CH1 coupling. . . . . . . . . . . . . . . . . . . . . . . . . DC
CH1 scale . . . . . . . . . . . . . . . . . . . . . . . . . . . 200 mV/div
CH1 input impedance . . . . . . . . . . . . . . . . . . 50
Horizontal
Sweep. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 500 ns/div or 400 ns/div
Trigger
Source. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CH1
Coupling . . . . . . . . . . . . . . . . . . . . . . . . . . . . DC
Slope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Positive
Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +100 mV
Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Auto