Tektronix AWG710B Ventilation Hood User Manual


 
Appendix B: Performance Verification (AWG710)
AWG710&AWG710B Arbitrary Waveform Generator User Manual B-101
3. Follow the substeps below to set the AWG710 Arbitrary Waveform Generator
controls and to select the sequence file:
a. Push UTILITY (front–panel)!System (bottom)!Factory Reset
(side)!OK (side).
b. Push SETUP (front–panel)!Run Mode (bottom)!Enhanced (side).
The AWG710 Arbitrary Waveform Generator is set to enhanced mode.
c. Load the AMP1.SEQ file.
Refer to Loading Files on page B-62 for file loading procedures.
4. Push the RUN button.
The LEDs above the RUN button is on.
5. Do the following substeps to set the AWG710 Arbitrary Waveform Generator
marker output high level setting:
a. Push VERTICAL MENU (front–panel)!Marker... (side)!Marker 1
High Level (side).
b. Push 3, ., 0 and ENTER keys in this order or turn the general purpose knob
to set the marker level to 3.0 V.
c. Verify that the marker 1 high level reading on the DMM screen falls within
2.75 V through 3.25 V.
d. Push the FORCE EVENT button.
6. Do the following substeps to set the AWG710 Arbitrary Waveform Generator
marker output low level setting:
a. Push Marker 1 Low Level side button.
b. Push -, 1, ., 1 and ENTER keys in this order or turn the general purpose
knob to set the marker level to -1.1 V.
c. Verify that the marker 1 low level reading on the DMM screen falls within
-1.255 V through -0.945 V.
d. Push the FORCE EVENT button.
7. Do the following substeps to change the connection to check the MARKER1
:
Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VDC
Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Auto
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Front