Tektronix AWG710B Ventilation Hood User Manual


 
Waveform Mixing Mode
AWG710&AWG710B Arbitrary Waveform Generator User Manual 3-247
Save/Restore Setup
You can save and restore the instrument output setup parameters. The setup
parameters on Extended Operation modes are saved together with the setting
information on AWG mode at a setting file.
Save/Restore operation is executed on the Save/Restore menu of the SETUP screen
in AWG mode. A saved setup file contains the setting information on both AWG
mode and each Extended Operation mode. Refer to The Save/Restore Menu on
page 3-48.
Operation Flow
When the AWG710&AWG710B is in AWG mode, change to Waveform Mixing
mode. Reference page 3-242.
1. Push Waveform/Sequence (bottomless... (side)!Load A File... or Load B
File... (side)
2. Push Previous Menu (side) to back to the Waveform/Sequence side menu.
3. Push Update (side) to select the Update mode.
Usually, select Auto. When generating a large mixed waveform, select
Manual.
4. Push Setting... (side) to specify the waveform mixing parameters.
Ratio of A, Ratio of B: Specify the ratio of Waveforms A and Waveform
B.
Marker: Specify which marker is used as a marker of a mixed waveform.
Wfm Len: Specify the length of the mixed waveform.
5. When Manual is selected in the Update menu, mixed waveform is updated
when Update Waveform (side) button is pushed.
6. When Manual is chosen in Update mode, change of a parameter will be
reflected if Update Waveform is pushed.
7. Set each item of Vertical, Horizontal, Run Mode, and the Trigger menu like the
time of AWG mode.
8. Set each parameter of the Vertical, Horizontal, Run Mode, and Trigger menu
is set up like the time of the usual AWG mode.
9. Push the RUN (front) button to turn on the RUN LED.
10. Push the CH1 OUT button and ALL MARKER OUT ON/OFF button to
output the signal that the corresponding output connector.