The Setup Menu Screen
3-54 AWG710&AWG710B Arbitrary Waveform Generator User Manual
Turning Marker Output
On and Off (AWG710B
only)
Push the ALL MARKER OUT ON/OFF button to connect or disconnect the
instrument marker output to the MKR1 and MKR2 connectors. When you push the
ALL MARKER OUT ON/OFF button, the ALL MARKER OUT ON/OFF LED
goes on and marker1 and marker2 signals are output from the MKR1 and MKR2
connector if the instrument is in the Running state. When you push this button
again, the marker signals output are disconnected and the marker signal outputs are
stopped, even if the instrument is in the Running state. If there is no waveform
loaded into a channel, you cannot turn that channel output on or off.
The ALL MARKER OUT ON/OFF LED automatically turns off when the
waveform data in that channel becomes invalid. For example, you attempt to load
an incorrect file, and the instrument deletes the current waveform from memory.
.
CAUTION. Turn OFF the ALL MARKER OUT ON/OFF button before
connecting/disconnecting a marker cable or turning on/off the power of DUT, in
order to prevent the damage to a marker circuit.
NOTE. The OUTPUT button and the ALL MARKER OUT ON/OFF button are
automatically turned off, when the mode of Extended Operation is changed, the
Connecting menu of the Synchronous Operation mode is pushed, the Trigger
Timing Cal of the Synchronous Operation mode is performed, and when Diag of
the UTILITY menu is performed.