Appendix B: Performance Verification (AWG710B)
B-42 AWG710&AWG710B Arbitrary Waveform Generator User Manual
b. Verify that the oscilloscope displays the waveform shown in Figure B-26
and that the waveform cycle is the half of Figure B-21.
Figure B-26: Waveform output when SW6 is closed
c. Open SW6 of the ground closure to degenerate the event signal.
d. Verify that the oscilloscope displays the waveform in Figure B-21.
12. Check the EVENT IN connector pin 5 input:
a. Close the SW7 of the ground closure to generate an event signal on the
EVENT IN connector pin 5.
b. Verify that the oscilloscope displays the waveform shown in Figure B-27.
Figure B-27: Waveform output when SW7 is closed
c. Open SW7 of the ground closure to degenerate the event signal.