Tektronix AWG710 Ventilation Hood User Manual


 
The UTILITY Window
3-174 AWG710&AWG710B Arbitrary Waveform Generator User Manual
Do the following steps to execute the calibration:
1. Push the RUN button to turn the output off if a waveform is being output.
The RUN LED is off.
2. Push UTILITY (front–panel)!Diag (bottom)!Execute Calibration (side).
The internal calibration routine runs immediately and requires up to 15 seconds
to complete.
The status message box appears when calibration has been terminated. See Figure
3-60.
Figure 3-60: Status message box (except option02)
Pass is displayed in the message box if the calibration successfully terminates. Fail
is displayed if calibration encounters a problem.
Push the OK side button or CLEAR MENU button to erase the status message box
and return to the screen shown in Figure 3-59.
Power-on Diagnostics
At power on, a limited set of hardware tests for all the test categories are performed
and the results are displayed on the screen. The instrument displays the following
message when an error is detected:
Press any key to go to the SETUP menu screen.