Appendix B: Performance Verification (AWG710)
B-72 AWG710&AWG710B Arbitrary Waveform Generator User Manual
Amplitude and Offset Accuracy Tests (Normal Out), (except option 02)
These procedures check the accuracy of the amplitude and offset outputs of the
AWG710 Arbitrary Waveform Generator.
Check Amplitude
Accuracy
Do the following steps to install the test hookup and set the test equipment controls:
1. Use a 50 BNC coaxial cable, a SMA(Ma)–BNC(Fe) adapter, a 50
precision terminator, and a BNC–to–dual banana adapter to connect the
AWG710 Arbitrary Waveform Generator CH1 output to the DMM input
connector (see Figure B-43).
NOTE. The amplitude and offset accuracy checks are structured as a continuous
test. The next test uses the control settings from the previous test and uses the next
step in the sequence file.
NOTE. When you output signal from the CH1 or CH1 OUTPUT, check that the
other OUTPUT ( CH1
or CH1 ) LED is off.
If the other OUTPUT LED is on, push the CH1
or CH1 OUT button to turn off the
output.
Equipment
required
A 50 BNC coaxial cable, a 50 precision terminator, a
SMA(Ma)–BNC(Fe) adapters, BNC (female)–to–dual banana adapter,
and a digital multimeter (DMM).
Prerequisites
The AWG710 Arbitrary Waveform Generator must meet the prerequisites
listed on page B-61.