Tektronix AWG710 Ventilation Hood User Manual


 
Appendix B: Performance Verification (AWG710)
B-82 AWG710&AWG710B Arbitrary Waveform Generator User Manual
4. Push SETUP (front–panel)!Run Mode (bottom)!Enhanced (side).
The AWG710 Arbitrary Waveform Generator is set to enhanced mode.
5. Load the AMP2.SEQ file.
Refer to Loading Files on page B-62 for file loading procedures.
6. Push the RUN and CH1 output buttons.
The LEDs above the RUN button and CH1 output connector are on.
7. Do the following substeps to check the option02 amplitude accuracy of 1 V
amplitude setting:
a. Write the DMM reading as a positive voltage.
b. Push the FORCE EVENT button.
c. Write the DMM reading as a negative voltage.
d. Verify that the positive minus negative voltages fall within
1.0 V ± 0.022 V.
e. Push VERTICAL MENU (front–panel)!Amplitude (side).
f. Push 0, . , 5 and ENTER keys in this order or turn the general purpose
knob to set the amplitude to 0.500 V.
8. Do the following to check the option02 amplitude accuracy of 500 mV
amplitude setting:
a. Push the FORCE EVENT button twice.
b. Write the DMM reading as a positive voltage.
c. Push the FORCE EVENT button.
d. Write the DMM reading as a negative voltage.
e. Verify that the positive minus negative voltages fall within
500 mV ± 12 mV.
9. Follow the substeps below to check the DC offset:
a. Push the FORCE EVENT button.
b. Verify that the reading on the DMM falls within 0 V ±10 mV.
10. Do the following substeps to change the connection to check the CH1
:
a. Push the CH1 output button.
b. Disconnect the adapter, terminator and cable from the CH1 output
connector.