Synchronous Operation Mode (AWG710B only)
AWG710&AWG710B Arbitrary Waveform Generator User Manual 3-257
Save/Restore Setup
You can save and restore the instrument output setup parameters. The setup
parameters on Extended Operation modes are saved together with the setting
information on AWG mode at a setting file.
Save/Restore operation is executed on the Save/Restore menu of the SETUP screen
in AWG mode. A saved setup file contains the setting information on both AWG
mode and each Extended Operation mode. Refer to The Save/Restore Menu on
page 3-48.
Operation Flow
Two AWG710Bs are set as Sync-Master and Sync-Slave in the following
procedure, and synchronous operation is performed.
1. In the power off state, connect two LAN cables, four clock signal cables and
four trigger signal cables
2. Push the front panel On/Standby switch to power on the instrument.
3. Check and write the IP address of slave
4. Push SETUP (front–panel)!Extended Operation (bottom)!
Sync - Slave... (side) button of the slave AWG710B.
5. Push SETUP (front–panel)!Extended Operation (bottom)!
Sync - Master... (side) button of the master AWG710B.
6. Push Connect (bottom)!Edit Slave IP Address...(side) button of the master
AWG710B to register the IP address of the slave AWG710B.
7. If required, push Execute Ping... (side) button to check that Slave can be
recognized on a network.
8. Push Connect to Slave (side) button to connect to the slave AWG710B.
9. Specify the files of the waveform to output and set the parameters of the
Vertical menu by Master and Slave.
10. The setting parameters about outputs, such as the Horizontal menu, are
performed by Master.
11. Push the CH1 OUT button and ALL MARKER OUT ON/OFF button to
output the signal that the corresponding output connector.
12. Push the RUN (front) button of the master AWG710B to start operation.