Appendix B: Performance Verification (AWG710B)
AWG710&AWG710B Arbitrary Waveform Generator User Manual B-51
a. Push UTILITY (front–panel)!System (bottom)!Factory Reset
(side)!OK (side).
b. Push SETUP (front–panel)!Run Mode (bottom)!Enhanced (side).
The AWG710B Arbitrary Waveform Generator is set to enhanced mode.
c. Load the AMP1.SEQ file.
Refer to Loading Files on page B-9 for file loading procedures.
4. Push the RUN button and ALL MARKER OUT ON/OFF button.
The LEDs above the RUN button and ALL MARKER OUT ON/OFF is on.
5. Do the following substeps to set the AWG710B Arbitrary Waveform
Generator marker output high level setting:
a. Push VERTICAL MENU (front–panel)!Marker... (side)!Marker 1
High Level (side).
b. Push 2, ., 4, 5 and ENTER keys in this order or turn the general purpose
knob to set the marker level to 2.45 V.
c. Verify that the marker 1 high level reading on the DMM screen falls within
2.2275 V through 2.6725 V.
d. Push the FORCE EVENT button.
6. Do the following substeps to set the AWG710B Arbitrary Waveform
Generator marker output low level setting:
a. Push Marker 1 Low Level side button.
b. Push -, 2, ., 0 and ENTER keys in this order or turn the general purpose
knob to set the marker level to -2.0 V.
c. Verify that the marker 1 low level reading on the DMM screen falls within
-2.2 V through -1.8 V.
d. Push the FORCE EVENT button.
7. Do the following substeps to change the connection to check the MARKER1
:
a. Disconnect the 50 precision terminator from the MARKER1
OUT
connector.
b. Disconnect the cable (with SMA(Ma)-BNC(Fe) adapter and BNC 50
terminator) from MARKER 1 OUT connector.
c. Connect the 50 precision terminator to the MARKER 1 OUT (removed
in step a).
d. Connect the adapter, terminator, and the cable (removed in step b) to the
MARKER1
OUT connector.
8. Press Force Event where needed.