Appendix B: Performance Verification (AWG710)
AWG710&AWG710B Arbitrary Waveform Generator User Manual B-87
Trigger Input Tests
These procedures check the trigger level accuracy of the AWG710 Arbitrary
Waveform Generator.
Do the following steps to install the test hookup and set the test equipment controls:
1. Use a 50 SMA coaxial cable and an SMA(Fe)–BNC(Fe) adapter to connect
the AWG710 Arbitrary Waveform Generator CH1 output connector to the
oscilloscope CH1 input connector.
2. Use a BNC coaxial cable to connect the function generator to the BNC–T
adapter which goes to the Trigger Input on the AWG710. Connect a 50 BNC
coaxial cable to the BNC–T adapter. Connect the opposite end of the BNC
cable to the CH2 input on the oscilloscope.
Figure B-49: Trigger input initial test hookup
3. Set the oscilloscope controls as follows:
Equipment
required
Two 50 BNC coaxial cables, a 50 SMA coaxial cable, a
SMA(Fe)–BNC(Ma) adapter, a BNC–T (male to 2 females) adapter, a
function generator, and an oscilloscope (TDS700).
Prerequisites
The AWG710 Arbitrary Waveform Generator must meet the prerequisites
listed on page B-61.
50 BNC coaxial cable
Function Generator (AFG310)
AWG710 Arbitrary Waveform
Generator rear panel
Oscilloscope (TDS700)
50 SMA coaxial cable
50 BNC coaxial cable
BNC T
Adapter
CH1 output
Output
CH1 input CH2 input
SMA(Female)-BNC
(Male) adapter
Vertical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CH1 and CH2
CH1 and CH2 coupling . . . . . . . . . . . . . . . . . DC
CH1 scale . . . . . . . . . . . . . . . . . . . . . . . . . . . 500 mV/div
CH2 scale . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 V/div