Appendix B: Performance Verification (AWG710B)
B-52 AWG710&AWG710B Arbitrary Waveform Generator User Manual
9. Repeat steps 5 and 6 for the AWG710B Arbitrary Waveform Generator
MARKER1
OUT.
10. Repeat steps 5 through 8 for MARKER 2 OUT and MARKER2
OUT.
11. Disconnect the DMM.
Synchronous Operation Tests
These procedures check the signal path for the Synchronous Operation of the
AWG710B Arbitrary Waveform Generator.
Do the following steps to install the test hookup and set the test equipment controls:
1. Use four 50 Ω SMA coaxial cables to connect the T IN connector to the T
OUT1 connector, the T IN
connector to the T OUT1 connector, the C IN
connector to the C OUT1 connector, and the C IN
connector to the C OUT1
connector(see Figure B-34).
Figure B-34: Synchronous operation test hookup
2. Follow the substeps below to set the AWG710B Arbitrary Waveform
Generator controls and to select the sequence file:
a. Push UTILITY (front–panel)!System (bottom)!Factory Reset
(side)!OK (side).
Equipment
required
Four 50 Ω SMA C coaxial cables (P/N: 174-1427-00).
Prerequisites
The AWG710B Arbitrary Waveform Generator must meet the
prerequisites listed on page B-7.
AWG710B Arbitrary Waveform Generator
50Ω SMA coaxial cable
T IN - T OUT 1
T IN - T OUT 1
C IN - C OUT 1
C IN - C OUT 1