Appendix B: Performance Verification (AWG710)
AWG710&AWG710B Arbitrary Waveform Generator User Manual B-73
Figure B-43: Amplitude accuracy initial test hookup
2. Set the DMM controls as follows:
3. Follow the substeps below to set the AWG710 Arbitrary Waveform Generator
controls and to select the sequence file:
a. Push UTILITY (front–panel)!System (bottom)!Factory Reset
(side)!OK (side).
b. Push SETUP (front–panel)!Run Mode (bottom)!Enhanced (side).
The AWG710 Arbitrary Waveform Generator is set to enhanced mode.
c. Load the AMP1.SEQ file.
Refer to Loading Files on page B-62 for file loading procedures.
4. Push the RUN and CH1 output buttons.
The LEDs above the RUN button and CH1 output connector are on.
Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VDC
Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Auto
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Front
AWG710 Arbitrary Waveform Generator
DMM
50 Ω BNC coaxial cable
CH1 output connector
+ SMA(Ma)-BNC(Fe) adapter
+ BNC 50 Ω precision terminator
+ 50 Ω BNC coaxial cable
Input connector
+ BNC-to-dual banana adapter
+ 50 Ω BNC coaxial cable