The filename of the power sensor data file entered could not be found on the currently
specified system mass storage file location. Check the filename and the system mass storage
file location.
Insufficient equip. to do test <test number> : <test name>
The required HP-IB-controlled test equipment for the test indicated is not available. Press
any key and choose another test.
No HP-IB address listed for DUT
Program attempted to address the spectrum analyzer (DUT), but no HP-IB address was
listed for it. Press any key and enter an address for the spectrum analyzer.
Non-numeric entry other than S entered, or frequency <=O
When entering a frequency of a frequency/Cal-Factor pair to be added, edited, or deleted, the
entry must either be a number greater than 0, or
S
to store the current data.
No sensor file found for <sensor model> S/N <sensor serial
#>
A power sensor data file for the indicated power sensor could not be found on the currently
specified system mass storage file location specifier. Check the sensor model and serial
numbers and the system mass storage file location.
Power meter reads <value>
dBm
The power meter has read a value far exceeding the specification of the CAL OUTPUT
amplitude; check that power sensor is connected to CAL OUTPUT and press any key. The
power will be read once more and assumed to be valid.
Select code <value> does not currently support HP-IB operations
The address just entered specified a select code which is not an HP-IB interface. Check the
address entered and the select code of the appropriate interface.
Sensor serial number must be from 1 to 99999
The power sensor serial number entered was not in the range indicated. Enter the serial
number correctly.
System mass storage file location catalog cannot be read
Program attempted to read the catalog of the system mass storage file location. Check the
msus of the system mass storage file location.
This program does not support the current DUT
The frequency response adjustment software supports only the HP 85623. Refer to the
Spectrum Analyzers section in “Getting Started,” earlier in this adjustment.
Frequency Response Adjustment Software 3-17