Figure 12-3. A7 First LO Distribution Amplifier Drive
Control
Latch
for
Band-Switch
Driver
Refer to function block P on Al4 Frequency Control Schematic Diagram (sheet 3 of 5) in the
HP 8560 E-Series Spectrum Analyzer Component Level Information.
1. Connect the positive lead of a DVM to A14U417 pin 14 and the negative lead to A14J18
pin 6.
2. Set the HP 85623 to the following settings:
Center frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
..300MHz
Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OHz
3. The voltage should measure approximately 0 Vdc (TTL low).
4. Set the HP 85623 center frequency to 3 GHz.
5. The voltage should measure approximately
f5
Vdc (TTL high).
12-12 RF Section