Procedure 14. W3 Line Switch Cable
Cl
1
(6 PLACES)
(6 PLACES)
-
(6 PLACES)
sml13e
Figure 4-27. Side Frame Mounting Screws
7. On the top side of the spectrum analyzer, redress W3.
8. Connect W3 to A6J2. Dress W3 into the slotted opening in the deck.
9. Connect
AlAlWl
to A3J602.
10. Secure the power-supply cover shield to the power supply using three
flathead
screws.
One end of the cover fits into a slot provided in the rear frame assembly. Ensure that the
extended portion of the cover shield is seated in the shield wall groove. See Figure 4-12.
11. Place W3 and the other cable assemblies between the CRT assembly and the power
supply cover so the W9 wires are below the surface of the power-supply cover.
12. Fold up the A2, A3, A4, and A5 assemblies into the spectrum analyzer as described in
“Procedure 5. A2, A3, A4, and A5 Assemblies Replacement,” steps 5 through 10.
13. Fold up Al4 and Al5 assemblies as described in “Procedure 9. Al4 and Al5 Assemblies
Replacement,” steps 3 through 5.
14. Replace the spectrum analyzer cover assembly.
15. Connect the line-power cord and switch the spectrum analyzer power on. If the spectrum
analyzer does not operate properly, turn off the spectrum analyzer power, disconnect the
line cord, and recheck the spectrum analyzer.
4-54 Assembly Replacement