Unable to load CONDITIONS file from listed system file location
Program attempted to load the CONDITIONS file from the listed system mass storage file
location. Check the msus of the System mass storage file location and the presence of the
CONDITIONS file.
Unable to load data from <sensor filename>
Program found power sensor data file, but could not read the data from the file. Use the
Sensor Utilities to delete the file and enter new data.
Unable to obtain catalog from <system mass storage file location>
The system could not verify that the system mass storage file location entered was available.
Check the msus of the system mass storage file location.
Unable to reach power level of <value>
dBm
Program was unable to set the source amplitude for a desired power meter reading. Check the
test setup.
WARNING: CONDITIONS file from system file location is not compatible!
CONDITIONS files from Operation Verification programs for the HP 8560 A/B-Series, or
HP 8560 E-Series analyzers are not compatible with the frequency response adjustment
software. Put the frequency response adjustment software in a different directory than the
Operation Verification programs, if possible. Default CONDITIONS data will be used.
Change the mass storage file location to a different location than either of the operation
verification programs. Then, modify the conditions data as necessary and press STORE
CONE3
.
Error
Messages
Beginning
with
Numeric
Characters
6481A Sensor cal data minimum frequency not
<=
50 MHz
Program requires the HP 8481A power sensor to have a Cal Factor at or below 50 MHz. Use
Sensor Utilities to add a Cal Factor at or below 50 MHz.
8481A Sensor cal data maximum frequency not
>=
300 MHz
Program requires the HP 8481A power sensor to have a Cal Factor at or above 300 MHz. Use
Sensor Utilities to add a Cal Factor at or above 300 MHz.
8485A Sensor cal data minimum frequency not = 10 MHz
Program requires the HP 8485A power sensor to have a Cal Factor at 10 MHz. Use Sensor
Utilities to add a Cal Factor at 10 MHz.
8485A Sensor cal data maximum frequency not
>=
26.5
GHz
Program requires the HP 8485A power sensor to have a Cal Factor at or above 26.5 GHz. Use
Sensor Utilities to add a Cal Factor at or above 26.5 GHz.
3-18 Frequency Response Adjustment Software